IWORX Data Acquisition System
|

Show images gallery
|
The iWorx 118 is a fast 100 Khz, high-resolution data acquisition system suitable for most data recording studies in research laboratories such as organ bath, ecg, pressure, etc .
|
- More information -
|
- Documents -
|
- Contact Bioseb -
|
Users
Technical specifications
Other products / accessory
- INSERM Faculté de médecine Lille, France
- ESCPI Paris, France
- Ecole Nationale Vétérinaire Alfort, France
- Faculté des sciences d'Orléans Orleans, France
- Service de santé des armées Grenoble, France
- Faculté des sciences de Toulouse Toulouse, France
|
Download PDF
|
|
This page is available in following languages:
|
|
|
The iWorx 118 is a very fast, high-resolution data acquisition system suitable for most data recording studies in the research laboratory. It offers 8 analog input channels, 8 digital outputs, 4 digital inputs and 2 DACs. The hardware is connected to PC-compatible computers via the popular USB interface (native construction) so setup is PLUG-AND-PLAY EASY !
The very small size of the unit ix-118 enable researcher to use very little bench space. Connector are BNC standards to facilitate interconnections with other devices.
System Requirements
The IX 118 requires for its SOFTWARE LABSCRIBE a minimum Pentium III processor running Windows 98, 2000 or XP with at least 128 megabytes of RAM, and at least 100 Mo of free space on the hard disk drive. An Internet connection may be required for EASY FREE UPGRADES of the software.
Analog Outputs
The IX 118 offers 2 +/- 10V Analog Outputs (DACs). Each of them is independently programmable and can be synchronized to work together. Basic stimulus parameters for each DAC, such as pulse width, frequency and amplitude, can be changed on-the-fly using handy controls located in the LabScribe software toolbar. Standard protocols include pulse, train, and step waveforms. Each standard protocol allows the quiescent state to be a holding voltage, thereby making it ideal for VOLTAGE CLAMPING applications.
Resolution and Noise
The iWorx 118 employs a 16-bit Analog/Digital converter to sample datas over a 10 Volt input range. Typical noise on any input is less than 1 millivolt (1 mV). This allows the recording of signals from 10 mV to 10 V without the need for additional external amplifier.
|
|
|
THE INFORMATION IN THIS WEB SITE AND IN LINKED PAGES AND DOCUMENTS IS PROVIDED "AS IS" AND DOES NOT CREATE ANY EXPRESS OR IMPLIED WARRANTY ABOUT BIOSEB OR ITS PRODUCTS OR SERVICES.
Information published on this Web Site as well as services, product specifications, availability and prices are subject to change without notice. BIOSEB may also make improvements and/or changes in the products and/or the programs described in this Web Site at any time without notice.
BIOSEB has made reasonable efforts to verify that the information in this Web site was accurate when first published. Such information may contain errors or omissions, however, and it is subject to change without notice. Bioseb does not undertake to update this information to include any such changes or to correct errors or omissions. Bioseb assumes no responsibility for any use of the information in this Web site or for any infringement of patents or other rights of third parties that may result. Certain information may be country-specific and may not apply in all countries.
|
Input
|
|
Number of Inputs |
8 single-ended BNC
|
|
Input Impedance |
1 MOhm
|
|
Input Range |
+10 V
|
|
Noise |
1 LSB (1 mV typical)
|
|
Gain |
x1
|
|
Digital Input Lines |
8 (4 dedicated and 4 programmable)
|
|
Output
|
|
Number of DACs |
2
|
|
DAC Resolution |
12 Bit
|
|
DAC Speed |
100k s/sec, independent of sample speed
|
|
DAC Output Range |
+10 V
|
|
DAC Modes |
Pulse, Train, Step, DC
|
|
DAC Performance
|
|
Pulse |
0.01
|
|
Width |
ms-6,500 ms (pulse mode)
|
|
Frequency |
0.2 hz-50k Hz (pulse mode)
|
|
Digital Output
|
|
Lines |
8 (4 dedicated and 4 programmable)
|
|
Digital Output Connector |
BNC A/D Converter
|
|
Sample Speed |
1 sample/sec -100k samples/sec
|
|
Resolution |
16 Bit
|
|
Interface |
USB
|
|
Other
|
|
Enclosure |
Aluminum (CE compliant)
|
|
Power |
20 V/220 V, 50/60 Hz
|
|
Software |
iLabScribe
|
|
Trigger Modes |
External Trigger, Threshold Trigger from Data, User Trigger
|
|
Display |
Real time, user definable screen time independent of sample rate, user definable units, AutoScale, full scale or user defined scale
|
|
Warranty and Upgrades |
The iWorx 118 hardware is protected with a three-year warranty. Software and firmware upgrades are free, and are automatically delivered to registered users via the Web. LabScribe software may be freely distributed. |
|
|
Model:
|
|
BIO-IX118
|
IWORX Data Acquisition System
|
|
Contact us
|
|
|
Accessories :
|
BIO-ETH-256 Two channels combinaison Bridge/ECG/EMG/EEG Amplifier
|
|
BIO-ETH 401 Low Noise four channels transducer amplifier |
|
|
|