CAPSTAR-100 CO2 Analyzer
|
|
|
Operating principle
The heart of the instrument is a miniature infrared analyzer with exceptional stability, low maintenance, and a linear output. The heated measurement cell prevents water condensation. The SENSORPOD, which contains the infrared measurement cell, attaches to the front panel by a 1.5m cable. This cell can be removed for cleaning, inspection, or replacement in minutes. The digital display shows CO2 concentration in percent or mmHg, and allows breath-by-breath peak end-tidal reading or continuous rapid response. Calibration is performed
with a single calibrated gas and room air.
An ideal capnometer for a wide range of applications
The reliability and ease of operation of the CAPSTAR-100 make it the perfect instrument for routine respiratory CO2 monitoring. It is the natural companion
to our SAR-830 series Small Animal Ventilators for monitoring respiratory status. With the AVS-1 Advanced Ventilator System, it can control end-tidal CO2. It can also be used to monitor unassisted animals. Gas sampling rates of less than 30ml/min can be used, making it possible to monitor animals as small as rats. The fast response time of the CAPSTAR-100 allows end-tidal response plateaus with respiratory rates of 120 breaths/min or greater. A range of accessories is available to ensure easy setup and convenient operation. An accessory pack containing sample tubing, moisture traps, and Nafion® moisture-absorbing sample tubes is included with the instrument.
Standard Features:
• End-tidal peak or continuous readings
• Low sample flow requirements
• Accurate and stable monitoring
• Linear output signal for recording
|
Applications:
• Respiratory monitoring
• Use with ventilated or unassisted animals
• Suitable for rats and larger animals
• CO2 control with AVS-1 Ventilator
|
|
|
|
THE INFORMATION IN THIS WEB SITE AND IN LINKED PAGES AND DOCUMENTS IS PROVIDED "AS IS" AND DOES NOT CREATE ANY EXPRESS OR IMPLIED WARRANTY ABOUT BIOSEB OR ITS PRODUCTS OR SERVICES.
Information published on this Web Site as well as services, product specifications, availability and prices are subject to change without notice. BIOSEB may also make improvements and/or changes in the products and/or the programs described in this Web Site at any time without notice.
BIOSEB has made reasonable efforts to verify that the information in this Web site was accurate when first published. Such information may contain errors or omissions, however, and it is subject to change without notice. Bioseb does not undertake to update this information to include any such changes or to correct errors or omissions. Bioseb assumes no responsibility for any use of the information in this Web site or for any infringement of patents or other rights of third parties that may result. Certain information may be country-specific and may not apply in all countries.
|
Analyzer type |
single beam infrared
|
|
Sample cell materials |
sapphire and stainless steel
|
|
Sample inlet connector |
0.062” OD brass tube
|
|
Recommended sample tubing size |
0.040" - 0.062" ID Tygon™
|
|
Measurement range |
0-10% CO2
|
|
Analog output scaling |
100mV/% CO2 (0-1V)
|
|
Linearity |
0.1% CO2
|
|
Repeatability |
0.1% CO2
|
|
Response time (T10 - T90) |
100mS at 100 ml/min sampling
|
|
Response time (T10 - T90) |
130mS at 50 ml/min sampling
|
|
Response time (T10 - T90) |
250mS at 30 ml/min sampling
|
|
Zero stability |
0.1% (8 hours), 0.2% (24 hours)
|
|
Operating temperature range |
5 - 40
|
|
Optical bench temperature |
48
|
|
Warm-up time |
5 min to 0.2%, 10 min to 0.1%
|
|
Sample rate flowmeter |
10 - 50 ml/min
|
|
Maximum sample cell pressure |
+5 psig
|
|
Interference effects: vaporized anesthetic agents |
negligible
|
|
Interference effects: 50% N2O |
0.1% at 0% CO2, 0.6% at 5% CO2 (uncompensated)
|
|
Nitrous oxide compensation (internal jumper for 25, 50, 75%) |
0.1% at 5% CO2
|
|
Controls |
Zero, Span (10-turn dials)
|
|
Electrical requirements |
120VAC/220VAC switchable, 35VA
|
|
Dimensions (main unit) |
9W x 5H x 9D in., 23W x 13H x 23D cm
|
|
Dimensions (SENSORPOD) |
3W x 2.5H x 3D in., 7.6W x 6.4H x 7.6D cm
|
|
Weight |
10 lbs (4.5kg) |
|
|
Model:
|
|
|
|
Linked products:
|
|
|