System 2000 Instrumentation
|
|
|
AC-916 Bioamplifier Plug-in
The AC-916 Bioamplifier is a SYSTEM 2000 plug-in module for EEG, ECG, EMG, and
other low-level biopotential signals. It features exceptionally low noise, a wide gain range, and sharp cutoff bandpass filters. Up to sixteen AC-916 modules can be installed in one CC-2000 card-cage for rack mounting.
The AC-916 features an instrumentation amplifier differential front-end, for the lowest noise and best common-mode rejection. Gain is adjustable from 20X - 100,000X in twelve steps, making it possible to resolve signals as small as a few microvolts in amplitude.
The individual channel inputs and outputs are connected to multi-pin connectors on the rear panel of the CC-2000 card-cage for convenient cable connections. Pre-wired input cables are available which simplify input electrode connections. The CC-2000 card-cage output connector is directly compatible with the PC-based WinDaq 16-channel data acquisition system, which is available as an option.
Note: Not to be used for human life support applications.
CC-2000 Sixteen Module Card-Cage
This is the rackmount or benchtop card-cage that is the foundation of any SYSTEM 2000 setup. It fits into a standard 19" equipment rack, and occupies 3U (5.25") height. The card-cage holds up to sixteen instrumentation modules of your choice. DC power for the modules is introduced via a DIN connector accessible from the rear of the chassis.
Inputs are made via a 50-pin connector on the rear of the card-cage. Pre-wired input cables are available as an option. Channel outputs are provided by BNC jacks on the rear panel, and via a 37-pin D-SUB connector. This output connector is also directly compatible with the Dataq Instruments WinDAQ data acquisition system. A handsome molded desktop enclosure (Model DTE-1000) is available as an option.
Standard Features:
• Holds sixteen modules
• Shielded all-aluminum enclosure
• Standard CC-2000 comes with universal power supply
|
Options:
• DTE-1000 Desktop enclosure
• ICA-916 Input cable assembly (w/ pin jacks)
• ICA-916/I Isolation amplifier input cable
|
|
THE INFORMATION IN THIS WEB SITE AND IN LINKED PAGES AND DOCUMENTS IS PROVIDED "AS IS" AND DOES NOT CREATE ANY EXPRESS OR IMPLIED WARRANTY ABOUT BIOSEB OR ITS PRODUCTS OR SERVICES.
Information published on this Web Site as well as services, product specifications, availability and prices are subject to change without notice. BIOSEB may also make improvements and/or changes in the products and/or the programs described in this Web Site at any time without notice.
BIOSEB has made reasonable efforts to verify that the information in this Web site was accurate when first published. Such information may contain errors or omissions, however, and it is subject to change without notice. Bioseb does not undertake to update this information to include any such changes or to correct errors or omissions. Bioseb assumes no responsibility for any use of the information in this Web site or for any infringement of patents or other rights of third parties that may result. Certain information may be country-specific and may not apply in all countries.
|
Input type |
Differential or single-ended, switchable
|
|
Input range for full-scale output |
+100mV - 0.5V
|
|
Input impedance |
>10,000MW
|
|
Noise, referred to input |
<5mV P-P, wideband
|
|
Common Mode Rejection |
>95dB
|
|
Gain range |
20X - 100,000X, 12-positions
|
|
Low-pass filter |
10Hz - 50kHz, 6-positions, -12dB/octave
|
|
High-pass filter |
0.1Hz - 300Hz, 6-positions, -12dB/octave
|
|
Power requirements |
+12VDC @20mA
|
|
Dimensions |
0.98W x 5.25H inches (25W x 133H mm) |
|
|
Model:
|
|
SYSTEM-2000
|
System 2000 Instrumentation
|
16 channels, low-noise |