CO2 monitoring is widely recognized as an important measure of the respiratory status of experimental animals. It is useful in setting ventilator parameters, and serves to gauge depth of anesthesia in unassisted, spontaneously breathing animals as well. The MicroCapStar extends this important technique to the realm of small experimental animals.

Operating principle

The MicroCapStar End-Tidal Carbon Dioxide Analyzer provides accurate end-tidal or continuous measurement of expired CO2 in animals as small as mice. It features very low sample flow requirements, rapid response time, and long-term stability. Respiratory rate (RR) is computed using the excursions of the CO2 waveform. The CO2 and RR measurements, as well as a trend plot of the end-tidal values, are displayed on the graphics LCD screen.

The heart of the MicroCapStar is a new temperature-controlled infrared CO2 sensor with digital output. Low sample flow and rapid response is achieved with a carrier gas system employing digitally-controlled active flow management. This technique precisely and automatically maintains the ratio of carrier flow to sample flow, which is essential for accurate measurements. The heated measurement cell prevents water condensation, even during long-term measurement sessions. The standard sample tubing set is 1m long, but lengths of up to 2.5m are possible.

Display panel
The front-panel display (see illustration) shows CO2 concentration (either instantaneous or ETCO2) in either percent or mmHg, and a 5-minute trend plot of ETCO2.
Calibration is performed with a single calibration gas and room air.
An adjustable ETCO2 alarm provides a warning when end-tidal values fall out of a user-adjustable preset range. All adjustments are performed digitally using a single knob. Built-in diagnostics monitoring warn of plugged sample tubing or other fault conditions.

Other products, options and accessories

The advanced features, reliability, and ease of operation of the MicroCapStar make it the perfect companion to our SAR-830 series Small Animal Ventilators for monitoring respiratory status. With the AVS-1 Advanced Ventilator System, it can be used to control end-tidal CO2. An accessory pack containing spare low-volume sample tubing and a variety of connectors and fittings is included with the instrument. A range of accessories is available to ensure easy setup and convenient operation. Windows-based monitoring software is included, which allows display of the measurements, and saving the data to a disk file.

Standard Features

• End-tidal peak or continuous readings
• Low sample flow requirements
• Accurate and stable monitoring
• Simple and easy one-gas calibration
• Linear output signal for recording
Applications

• Suitable for mice and other small animals
• Respiratory monitoring
• Use with ventilated or unassisted animals
• Verifies proper ventilator operation
• CO2 control using AVS-1 Ventilator

THE INFORMATION IN THIS WEB SITE AND IN LINKED PAGES AND DOCUMENTS IS PROVIDED "AS IS" AND DOES NOT CREATE ANY EXPRESS OR IMPLIED WARRANTY ABOUT BIOSEB OR ITS PRODUCTS OR SERVICES.

Information published on this Web Site as well as services, product specifications, availability and prices are subject to change without notice. BIOSEB may also make improvements and/or changes in the products and/or the programs described in this Web Site at any time without notice.

BIOSEB has made reasonable efforts to verify that the information in this Web site was accurate when first published. Such information may contain errors or omissions, however, and it is subject to change without notice. Bioseb does not undertake to update this information to include any such changes or to correct errors or omissions. Bioseb assumes no responsibility for any use of the information in this Web site or for any infringement of patents or other rights of third parties that may result. Certain information may be country-specific and may not apply in all countries.

Carbon dioxide measurement Single beam, non-dispersive infrared
Measurement range 0 - 9.9% (0 - 76.0mmHg) CO2
Accuracy 0.15% (1.1mmHg)
Resolution (internal and analog output) 0.01% (0.1mmHg)
Resolution (display) 0.1% (1.0mmHg)
Linearity 0.1% CO2
Repeatability 0.1% CO2
Response time (T10 - T90) 75mS at 70 ml/min through cell
Sample cell materials Sapphire and stainless steel
Analog output scaling 1.1V/% (0-10V)
Response time (T10 - T90) 150mS at 50 ml/min sampling
Zero stability 0.2% (8 hours), 0.3% (24 hours)
Interference effects: 50% N2O 0.1% at 0% CO2, 0.6% at 5% CO2 (uncomp.)
Maximum sample cell pressure +5 psig
Interference effects: vaporized anaesthetic agents Negligible
Operating temperature range 5-40° C
Optical bench temperature 48° C, controlled
Warm-up time 4 min to 0.2%, 10 min to 0.1%
ETCO2 trend display 5 minute graphical scrolling display
Sample flow (sample inlet - carrier flow out) 10ml/min nominal, adjustable from 5.0 - 20ml/min
Sample tubing connections (carrier flow, sample inlet) Luer female
Calibration controls Zero, Span
ETCO2 alarm adjustment range 1.0 - 9.9% (8 - 75mmHg)
Respiratory rate measurement range 5 - 200 breaths/min
Signal outputs (rear panel) BNC jacks
CO2 output scaling 1.0V / %CO2
Respiratory rate output scaling 0.05V / bpm
Alarm output 5V if alarm condition, 0V if no alarm
Serial data output format 2400 baud, 8 data, no parity, 1 stop bit
Electrical requirements 120VAC/220VAC switchable, 35VA
Dimensions 19W x 5.25H x 16D in., 49W x 13H x 41D cm
Weight 10 lbs. (4.5kg)

Model:
MicroCapstar
MicroCapStar
Contact us

Related products:
CAPSTAR-100
CAPSTAR-100 CO2 Analyzer
Contact us
OxyStar
OxyStar-100 O2 Monitor
Contact us
GSM-3
GSM-3 Gas Mixer
Contact us
GEMINI
End-tidal O2 and CO2 monitor
Complete with accessory pack, 110/220V Contact us
Print version

Bioseb - In Vivo Research Instruments
Phone worldwide : +33 442 344 360 - USA/Canada : (727) 521-1808
e-Mail : Worldwide: info@bioseb.com - USA/Canada: info-us@bioseb.com